A large number of certified reference materials (CRMs) are polished and mounted,
one by one, on the spark stand. The signals from these samples are recorded and
plotted against their certified concentrations.
![[Spark calibration]](images/spark1.gif)
Example calibration of Mn in Steel/Cast Iron
Copyright © Jobin-Yvon Emission 1999, with permission
A separate calibration is made for each type of material
(matrix) being analysed, for example, cast iron, aluminium, etc.
Software allows various types of corrections to improve the fit
between the measured signals and the certified concentrations, and hence the
accuracy of analysis, eg, spectral interferences, self-absorption.
Authors: Jean Charles Lefebvre, Jobin-Yvon Emission
and Richard Payling, Surface Analytical
First published on the web: 15 November 1999.